OECD TG 105
OECD TG 105 is the ‘Water Solubility’ standard, i.e. water solubility, which is listed under Section 1 (Physico-chemical properties) of the OECD Test Guidelines for the Testing of Chemicals.
Water solubility is a measure of the amount of a substance that can be dissolved in pure water at a specific temperature (usually 20 °C), expressed as mass/volume or mass/mass. OECD TG 105 is particularly suitable for substances with low and medium solubility. For highly soluble substances, other approaches may be required.
The basic principle of the test is to bring the test substance into equilibrium with water and to determine the concentration of the dissolved substance using a suitable analytical method. OECD TG 105 comprises two basic methods:
The shaking flask method:
- The test substance is shaken together with water in a sealed container
- We wait until equilibrium is reached (usually 24–48 hours)
Following phase separation, the concentration of the substance in the solution is determined analytically
Suitable for substances with medium and relatively high solubility
Column elution method:
- The test substance is applied to an inert carrier material and loaded into a column
- Water is passed through the column; the dissolved substance is collected in the eluate
- The concentration in the eluate is measured analytically
- Suitable for substances with low solubility; requires less sample material
The key parameters reported in the study are:
- Water solubility (mg/l or g/l)
- Measured temperature (°C)
- pH value (solubility may depend on the pH value)
- Method used (flask or column elution)
- Analytical measurement method and limit of detection
- Data on measurement uncertainty and repeatability
Why is this being done?
OECD TG 105 is being carried out for the following scientific and regulatory purposes:
- Substance characterisation: Water solubility is a fundamental physico-chemical property of a substance and is used for identification purposes
- Predicting environmental behaviour: Solubility provides an indication of the distribution, bioavailability, transport and fate of the substance in water
- Design of ecotoxicity tests: Solubility is a key factor in the selection of test concentrations and dosing
- Assessment of bioaccumulation potential: Low water solubility may be an indicator of a tendency to bioaccumulate
- Classification and labelling: Solubility is taken into account when classifying chemicals in terms of physical and environmental hazards
- Support with the preparation of authorisation dossiers: Used in REACH, biocidal product, plant protection and medicinal product safety assessments
- Quality control: Monitoring the consistency of raw materials, intermediate products and finished products in production processes
- Weighing up the evidence: a contribution to the holistic assessment of the substance in conjunction with other physicochemical, toxicological and environmental data
OECD TG 105 is not a standalone toxicity or hazard assessment. It is intended for physicochemical characterisation; the results must be interpreted in the context of other physicochemical properties, toxicological data and the exposure assessment.
As part of our nano-testing services, we carry out water solubility determinations in accordance with OECD TG 105, applying scientific rigour and drawing on our accredited laboratory facilities and quality management system. Our scope of services includes the following:
Method selection and validation
Selection of the appropriate method (bottle or column elution) based on the expected solubility range and the properties of the test substance
Validation of the selected method and verification using suitable reference substances; selection, calibration and determination of the limit of detection for the analytical measurement method (HPLC, GC, spectrophotometry, etc.)Assessment and reporting of measurement uncertainty; documentation of method-specific limitations and conditions of use
Please feel free to contact Nano-Test for information on the choice of methods, the analytical measurement strategy, sample reception conditions, the test schedule and the reporting procedures for your water solubility determinations in accordance with OECD TG 105.