OECD TG 113
OECD TG 113 is the standard ‘Screening Test for Thermal Stability and Stability in Air’, i.e. the screening test for thermal stability and stability in air, which is listed under Section 1 (Physicochemical Properties) of the OECD Test Guidelines for the Testing of Chemicals.
The aim of the test is to assess, in advance, the thermal stability of a substance when heated and its oxidative stability on contact with air. OECD TG 113 is a screening test; rather than determining precise thermodynamic or kinetic parameters, it aims to establish quickly and reliably whether the substance decomposes in a hazardous manner.
Thermal stability is a measure of the extent to which a substance retains its chemical structure when heated. Stability in air, on the other hand, describes the substance’s resistance to oxidation or other air-related reactions in the presence of oxygen. The test enables these two properties to be assessed together.
OECD TG 113 does not prescribe any specific method, but permits the use of suitable and validated thermal analysis methods. Common methods include:
Differential Scanning Calorimetry (DSC):
- The specimen is heated according to a controlled temperature programme, and the change in heat flux is measured
- Exothermic and endothermic transitions, the decomposition onset temperature and the peak temperature are determined
- To determine oxidative stability, the sample is analysed in air or an oxygen atmosphere; to determine thermal stability, it is analysed in an inert atmosphere (nitrogen, argon), and the results are compared
Thermogravimetric analysis (TGA):
- The change in the mass of the sample is measured as a function of temperature
- The initial temperature at which mass loss begins, the decomposition stages and the residual amount are determined
- Comparing the measurements taken in air and in an inert atmosphere enables the oxidative decomposition to be assessed
Differential thermal analysis (DTA):
- The temperature difference between the sample and the reference material is measured
- Used to detect exothermic and endothermic processes
- Provides similar information to the DSC; may be the preferred option depending on the device’s features
Simultaneous DSC/TGA (STA):
- Simultaneous measurement of thermal and gravimetric changes in the same sample
- Enables the interpretation of decomposition processes on the basis of changes in heat and mass
Tests using a closed or open crucible:
- Sealed crucibles for assessing pressure build-up and volatile products
- Open crucibles for assessing the effect of oxygen access
The key parameters reported in the study are:
Initial decomposition temperature (°C) and peak temperature
Oxidation onset temperature (measured in air)
Mass loss (%) and temperature at the start of mass loss; identification of exothermic/endothermic processes; atmosphere used (air, oxygen, nitrogen, argon); heating rate, sample quantity and vessel type; data on measurement uncertainty and repeatability
Why is this being done?
OECD TG 113 is being carried out for the following scientific and regulatory purposes:
Determination of thermal stability: establishing whether the substance decomposes at storage, transport and processing temperatures
Assessment of oxidative stability: determination of the substance’s tendency to oxidise or react spontaneously on contact with air
Hazard identification: Early detection of thermally unstable substances or those that react with air (e.g. self-heating substances, oxidising agents)
Classification and labelling: Assessment of thermal stability and the potential for spontaneous reactions in the context of transport regulations (UN/ADR) and chemical classification (CLP/GHS)
Establishing safe storage and transport conditions: providing information on the maximum safe processing and storage temperatures
Process safety: Preliminary assessment of thermal risks in chemical processes
Input for further tests: Data on thermal stability influence the design and conditions under which other physicochemical and toxicological tests are carried out
- Quality control: Monitoring the consistency of the material during the manufacturing and storage processes
- ‘Weight-of-evidence’ approach: a contribution to the holistic assessment of the substance in conjunction with other physicochemical data and safety assessments
OECD TG 113 does not, in itself, constitute a toxicity or comprehensive hazard assessment. It provides physicochemical information as part of a screening process; the results should be interpreted in conjunction with other thermal analysis data, reactivity tests and safety assessments.
As part of our nano-testing services, we carry out thermal stability and air stability screening in accordance with OECD TG 113, applying scientific rigour within the framework of our accredited laboratory infrastructure and quality management system. Our scope of services includes the following:
Method selection and validation
- Selection of the appropriate method (DSC, TGA, DTA, STA) in accordance with the properties of the test substance and regulatory requirements
Choice of atmosphere: inert atmosphere (nitrogen/argon) for thermal stability; air/oxygen for oxidative stability
Validation of the selected method and verification using suitable reference substances (e.g. indium, zinc, tin)
Calibration of temperature and heat flux; evaluation and reporting of measurement uncertainty
Documentation of method-specific limitations and conditions of use
Please feel free to contact Nano-Test for information on the selection of methods and atmospheres, sample acceptance criteria, the test plan and the reporting procedures for your thermal stability and air stability tests in accordance with OECD TG 113.

