OECD TG 101
OECD TG 101, OECD Test Guidelines for the Testing of Chemicals, Section 1 (Physico-chemical Properties) in the context of ‘UV-VIS absorption spectra’, i.e. UV-VIS absorption spectra. It was first adopted in 1981 and updated in 1995. It is not a test for health effects or ecotoxicity, but rather a test of physicochemical properties.
The aim of the test is to determine the absorption spectrum of a substance in the ultraviolet (UV) and visible (VIS) regions. The absorption spectrum shows at which wavelengths and to what extent the substance absorbs light; this information provides fundamental insights into the chromophore structure, the electronic transitions and the substance’s potential to interact with light.
The basic principle of the test is as follows:
- The test substance is dissolved in a suitable solvent (usually water); if solubility is low, suitable organic solvents (e.g. methanol, ethanol, hexane) may be used.
- Die Extinktion der Lösung wird mit einem UV-VIS-Spektralphotometer im Wellenlängenbereich von in der Regel 200–800 nm gemessen.
- The spectrum is recorded as an absorption-wavelength curve; the wavelengths at which absorption is greatest (λmax) and the corresponding absorption values are determined.
- Using Lambert–Beer’s law (A = ε·c·l), the molar absorption coefficient (ε) can be calculated when the concentration and the optical path length are known.
- If necessary, the spectrum is measured again at different pH values (for ionisable substances, the values for λmax and ε may vary depending on the pH).
The key parameters reported in the study are as follows:
- Absorption maxima (λmax, nm) and the corresponding absorption values
- Molar absorption coefficient (ε, L·mol⁻¹·cm⁻¹) — where calculable
- Measurement wavelength range (nm)
- Solvent used and solution concentration
- Optical path length (cuvette thickness, cm)
- Measured temperature (°C) and pH value (where applicable)
- Spectral curve and, where applicable, characteristic absorption bands
- Data on instrument calibration and measurement uncertainty
What is it for?
OECD TG 101 is being carried out for the following scientific and regulatory purposes:
- Substance identification and characterisation: The UV-VIS spectrum is used to identify a substance and to determine its structural properties
- Determination of the chromophore structure: Absorption bands provide information about conjugated systems and functional groups within the molecule
- Prediction of photochemical behaviour: Substances that exhibit absorption in the UV-VIS region may have the potential for photodegradation or phototoxicity when exposed to sunlight
- Preliminary assessment of phototoxicity: Light absorption is used to assess the need for phototoxicity testing (e.g. OECD TG 432)
- Prediction of environmental fate: The absorption spectrum provides an indication of the substance’s potential to be degraded in the environment through photolysis
- Development of analytical methods: This forms the basis for wavelength selection in analytical methods based on UV-VIS detection (HPLC-UV, etc.)
- Support with the preparation of authorisation dossiers: Used in REACH, biocidal product, plant protection and medicinal product safety assessments
- Quality control: Monitoring the consistency of the spectral properties of materials in production processes
- Weighting approach: a contribution to the holistic assessment of the substance in conjunction with other physicochemical, toxicological and environmental data
OECD TG 101 is not a standalone toxicity or hazard assessment. It is intended for physicochemical characterisation; the results must be interpreted in conjunction with other physicochemical properties, photochemical data, toxicological data and the exposure assessment.
As part of our nano-testing services, we carry out UV-VIS absorption spectrum measurements in accordance with OECD TG 101, applying scientific rigour and utilising our accredited laboratory facilities and quality management system. Our scope of services includes the following:
Method selection and validation
- Selection and justification of the appropriate solvent (water or organic solvent) based on the solubility and properties of the test substance
- Determination of the measurement wavelength range and the concentration
- Calibration of the spectrophotometer (wavelength and absorbance accuracy, using reference substances) and performance monitoring
- Validation of the method: linearity, accuracy, repeatability and inter-repeatability
- Evaluation and reporting of measurement uncertainty
- Documentation of method-specific limitations and conditions of use
Please feel free to contact Nano-Test for information on solvent selection, measurement strategy, sample reception conditions, the test plan and reporting procedures for your UV-VIS absorption spectrometry measurements in accordance with OECD TG 101.